Single Phase Power Analyser
May 30, 2012 by SMVIT PRESS
Filed under Tender
Item No. 7: Single Phase Power Analyser
Quantity: 2
Specifications:
Basic measurement: V,A,W,VA,VAr,PF,Apk, Vpk, Acf,
Frequency Integrator: Wh, Vah, VArh, Ah, average PF, Timer
Inrush: 0-200Apk, Displays peak current during switch on.
Ballast mode: For measurement on electronic lighting ballasts, Ballast CT
Voltage: Range 2V rms to 1000Vpk (8 ranges)
Accuracy: 0.1% rdg +/-0.1% range +/- 0.1% per KHz+/- 10mV.
Overload withstand: 60A rms for 1 sec.
Watt: Range 0 to 1999GWAccuracy 0.2% rdg +/- 0.2% Range +/-5mW +/-0.3% per KHz.
Apparent power: Range 0 to 1999GVA, Accuracy 0.2% rdg +/- 0.2% range +/-5mVA
+/- 0.3% per KHz.
Reactive power: Range 0 to 1999GVAR, Accuracy 0.2% rdg +/- 0.2% range +/-5mVAR
+/-0.3% per KHz.
Power factor: Range 0.0000 to +/- 1.0000, Accuracy +/-0.004 +/-0.2%rdg +/-0.2%
per KHz
Frequency: Range 5Hz to 50Khz, Accuracy 0.2%
Harmonics: V,A,W Range 0 to 50th harmonics, Bandwidth Fundamental 5Hz to 50kHz, Harmonics 10Hz to 250Khz.
THD: Range 0% to 999%
Crest Factor: Range 1.000 to 19.9999
Integrator: Range 1 min to 99Hrs 59 Mins
Bandwidth: DC and 5Hz to 250khz
Safety and EMC:IEC 61010, complies with CE directive
Line power input: Dual Range 90-120V, 195-265V, 48-65Hz
Interfaces: IEEE 488, RS232 I/O and Parallel printer port, chart recorders and alarms
Dimensions: Suitable for rack mounting